Comparison of receive inserts for non-human primate fMPI in a human brain scanner

2025/03/14 by Niebel, Frauke H., Barksdale, Alex C., Sliwiak, Monika +3
#Coil design #Functional magnetic particle imaging #Non-Human Primate

paper · doi:10.18416/ijmpi.2025.2503020

Abstract

The receive coil geometry significantly influences MPI detection sensitivity, a primary distinguishing characteristic when competing with other tracer-based imaging methods. Here, we compare four gradiometric receive inserts for non-human primate functional MPI to identify a coil geometry that maximizes sensitivity while accommodating the primate head. A cylindrical coil with a diameter of  14 cm provided the highest sensitivity while fitting the macaque. Smaller cross-section gradiometers offered better rejection of noise originating from the shift coil amplifiers.