2025/01/13 by Abramowicz, H., Soto, M. Almanza, Benhammou, Y. +21
#FOS: Physical sciences #High Energy Physics - Experiment (hep-ex) #Instrumentation and Detectors (physics.ins-det)
paper · doi:10.48550/arxiv.2501.07431
Two samples of silicon pad sensors and two samples of GaAs sensors are studied in an electron beam with 5 GeV energy from the DESY-II test-beam facility. The sizes of the silicon and GaAs sensors are about 9×9 cm2 and 5×8 cm2, respectively. The thickness is 500 micrometer for both the silicon and GaAs sensors. The pad size is about 5×5 mm2. The sensors are foreseen to be used in a compact electromagnetic sampling calorimeter. The readout of the pads is done by metal traces connected to the pads and the front-end ASICs at the edges of the sensors. For the silicon sensors, copper traces on a Kapton foil are connected to the sensor pads with conducting glue. The pads of the GaAs sensors are connected to bond-pads via aluminium traces on the sensor substrate. The readout is based on a dedicated front-end ASIC, called FLAME. Pre-processing of the raw data and deconvolution is performed with FPGAs. The whole system is orchestrated by a Trigger Logic Unit. Results are shown for the signal-to-noise ratio, the homogeneity of the response, edge effects on pads, and for signals due to the readout traces.