2022/10/10 by Neal Patwari, Di Huang, Patwari, Neal +3
Engineering · Environmental Science · Medicine · #Climate Change and Health Impacts #FOS: Electrical engineering #Non-Invasive Vital Sign Monitoring #Signal Processing (eess.SP) #Ultrasound in Clinical Applications #electronic engineering #information engineering
paper · pdf · doi:10.48550/arxiv.2210.04990
openalex publication_date 2022/10/10 · openalex created_date 2022/10/14 · openalex updated_date 2026/07/28
Studies have shown pulse oximeter measurements of blood oxygenation have statistical bias that is a function of race, which results in higher rates of occult hypoxemia, i.e., missed detection of dangerously low oxygenation, in patients of color. This paper further characterizes the statistical distribution of pulse ox measurements, showing they also have a higher variance for patients racialized as Black, compared to those racialized as white. We show that no single race-based correction factor will provide equal performance in the detection of hypoxemia. The results have implications for racially equitable pulse oximetry.