Prescanning Assembly Optimization Criteria for Computed Tomography

2023/09/29 by Mayank Goswami, Goswami, Mayank
Engineering · Medicine · #Advanced MRI Techniques and Applications #Advanced X-ray and CT Imaging #FOS: Electrical engineering #FOS: Physical sciences #Image and Video Processing (eess.IV) #Instrumentation and Detectors (physics.ins-det) #Medical Imaging Techniques and Applications #Medical Physics (physics.med-ph) #electronic engineering #information engineering

paper · pdf · doi:10.48550/arxiv.2309.17303

openalex publication_date 2023/09/29 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/28

Abstract

Computerized Tomography assembly and system configuration are optimized for enhanced invertibility in sparse data reconstruction. Assembly generating maximum principal components/condition number of weight matrix is designated as best configuration. The gamma CT system is used for testing. The unoptimized sample location placement with 7.7% variation results in a maximum 50% root mean square error, 16.5% loss of similarity index, and 40% scattering noise in the reconstructed image relative to the optimized sample location when the proposed criteria are used. The method can help to automate the CT assembly, resulting in relatively artifact-free recovery and reducing the iteration to figure out the best scanning configuration for a given sample size, thus saving time, dosage, and operational cost.