Calibration of the charge and energy loss per unit length of the MicroBooNE liquid argon time projection chamber using muons and protons

2019/07/31 by MicroBooNE collaboration, C. Adams, M. Alrashed +186 · 1 citation
Physics and Astronomy · #physics.ins-det #hep-ex

paper · pdf · doi:10.1088/1748-0221/15/03/p03022

published as JINST 15, P03022 (2020) · Accepted version

arxiv created 2020/02/25 · arxiv updated 2020/04/07

Abstract

We describe a method used to calibrate the position- and time-dependent response of the MicroBooNE liquid argon time projection chamber anode wires to ionization particle energy loss. The method makes use of crossing cosmic-ray muons to partially correct anode wire signals for multiple effects as a function of time and position, including cross-connected TPC wires, space charge effects, electron attachment to impurities, diffusion, and recombination. The overall energy scale is then determined using fully-contained beam-induced muons originating and stopping in the active region of the detector. Using this method, we obtain an absolute energy scale uncertainty of 2% in data. We use stopping protons to further refine the relation between the measured charge and the energy loss for highly-ionizing particles. This data-driven detector calibration improves both the measurement of total deposited energy and particle identification based on energy loss per unit length as a function of residual range. As an example, the proton selection efficiency is increased by 2% after detector calibration.