vix.ing · top · new · best · stats · spec

MRI for attenuation correction in PET: methods and challenges

2012/11/21 by Gudrun Wagenknecht, H. Kaiser, Felix M. Mottaghy +1 · 1 citation
Medicine · Engineering · #Medical Imaging Techniques and Applications #Advanced MRI Techniques and Applications #Advanced X-ray and CT Imaging

paper · pdf · doi:10.1007/s10334-012-0353-4

openalex publication_date 2012/11/21 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/31

Abstract

In current combined PET/MR systems, PET attenuation correction is based on MRI, since the small bore inside MRI systems and the strong magnetic field do not permit a rotating PET transmission source or a CT device to be integrated. Unlike CT measurements in PET/CT scanners, the MR signal is not directly correlated to tissue density and thus cannot be converted by a simple transformation of intensity values. Various approaches have been developed based on templates, atlas information, direct segmentation of T1-weighted MR images, or segmentation of images from special MR sequences. The advantages and disadvantages of these approaches as well as additional challenges will be discussed in this review.

Cited by

Related