Demountable coaxial gas-cooled current leads for MRI superconducting magnets

1992/01/01 by C.J. Zhang, Xingmei Zhou, Leihong Yang · 1 citation
Engineering · #Superconducting Materials and Applications #Particle Accelerators and Free-Electron Lasers #Advanced Thermodynamic Systems and Engines

paper · doi:10.1109/20.120038

openalex publication_date 1992/01/01 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/29

Abstract

To fabricate a demountable current lead for MRI (magnetic resonance imaging) superconducting magnets with less heat loss, the heat transfer equation of a gas-cooled current lead is introduced. The authors discuss the character of demountable coaxial gas-cooled current leads and make careful selection of lead materials. Leads of 250 A for MRI magnets were made and tested. It is concluded that these current leads are adaptable to all kinds of closed-loop superconducting magnets and very convenient to use. The lead is safe to use even when there is no helium gas glow in it.>