Attenuation length measurements of liquid scintillator with LabVIEW and reliability evaluation of the device

2013/05/07 by L. Gao, BX. Yu, YY. Ding +17 · 1 citation
Physics and Astronomy · #hep-ex #physics.ins-det

paper · pdf · doi:10.1088/1674-1137/37/7/076001

5 pages, 8 figures, 9 conferences

arxiv created 2013/05/07 · arxiv updated 2015/06/15

Abstract

The attenuation length measuring device was constructed by using oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested with a variety of ways, the test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.10 plus or minus 0.35 m where Gd-LS was heavily used in Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment.