vix.ing · top · new · best · stats · spec

Influence of surface characteristics on bone integration of titanium implants. A histomorphometric study in miniature pigs

1991/07/01 by D. Buser, Daniel Buser, Robert Schenk +7 · 33 citations
Dentistry · Engineering · Medicine · #Dental Implant Techniques and Outcomes #Bone Tissue Engineering Materials #Orthopaedic implants and arthroplasty

paper · doi:10.1002/jbm.820250708

Abstract

The purpose of the present study was to evaluate the influence of different surface characteristics on bone integration of titanium implants. Hollow-cylinder implants with six different surfaces were placed in the metaphyses of the tibia and femur in six miniature pigs. After 3 and 6 weeks, the implants with surrounding bone were removed and analyzed in undecalcified transverse sections. The histologic examination revealed direct bone-implant contact for all implants. However, the morphometric analyses demonstrated significant differences in the percentage of bone-implant contact, when measured in cancellous bone. Electropolished as well as the sandblasted and acid pickled (medium grit; HF/HNO3) implant surfaces had the lowest percentage of bone contact with mean values ranging between 20 and 25%. Sandblasted implants with a large grit and titanium plasma-sprayed implants demonstrated 30-40% mean bone contact. The highest extent of bone-implant interface was observed in sandblasted and acid attacked surfaces (large grit; HCl/H2SO4) with mean values of 50-60%, and hydroxylapatite (HA)-coated implants with 60-70%. However, the HA coating consistently revealed signs of resorption. It can be concluded that the extent of bone-implant interface is positively correlated with an increasing roughness of the implant surface.

Cited by

Related