2013/10/07 by Ryan Brown, Pippa Storey, Christian Geppert +6 · 1 citation
Medicine · #Advanced MRI Techniques and Applications #Cardiac Imaging and Diagnostics #MRI in cancer diagnosis
paper · pdf · doi:10.1002/jmri.24205
openalex publication_date 2013/10/07 · openalex created_date 2025/10/10 · openalex updated_date 2026/08/04
PURPOSE: To develop a bilateral coil and fat suppressed T1-weighted sequence for 7 Tesla (T) breast MRI. MATERIALS AND METHODS: A dual-solenoid coil and three-dimensional (3D) T1w gradient echo sequence with B1+ insensitive fat suppression (FS) were developed. T1w FS image quality was characterized through image uniformity and fat-water contrast measurements in 11 subjects. Signal-to-noise ratio (SNR) and flip angle maps were acquired to assess the coil performance. Bilateral contrast-enhanced and unilateral high resolution (0.6 mm isotropic, 6.5 min acquisition time) imaging highlighted the 7T SNR advantage. RESULTS: Reliable and effective FS and high image quality was observed in all subjects at 7T, indicating that the custom coil and pulse sequence were insensitive to high-field obstacles such as variable tissue loading. 7T and 3T image uniformity was similar (P=0.24), indicating adequate 7T B1+ uniformity. High 7T SNR and fat-water contrast enabled 0.6 mm isotropic imaging and visualization of a high level of fibroglandular tissue detail. CONCLUSION: 7T T1w FS bilateral breast imaging is feasible with a custom radiofrequency (RF) coil and pulse sequence. Similar image uniformity was achieved at 7T and 3T, despite different RF field behavior and variable coil-tissue interaction due to anatomic differences that might be expected to alter magnetic field patterns.