2025/01/01 by Luiz E. Luiz, Salviano Soares, António Valente +3 · 1 voice
Medicine · Engineering · #ECG Monitoring and Analysis #Wireless Body Area Networks #Wireless Power Transfer Systems
paper · pdf · doi:10.1016/j.csbj.2025.04.020
openalex publication_date 2025/01/01 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/31
PROBLEM: Portable ECG/sEMG acquisition systems for telemedicine often lack application flexibility (e.g., limited configurability, signal validation) and efficient wireless data handling. METHODOLOGY: A modular biosignal acquisition system with up to 8 channels, 24-bit resolution and configurable sampling (1-4 kHz) is proposed, featuring per-channel gain/source adjustments, internal MUX-based reference drive, and visual electrode integrity monitoring; Bluetooth® transmits data via a bit-wise packet structure (83.92% smaller than JSON, 7.28 times faster decoding with linear complexity based on input size). RESULTS: input-referred noise; harmonic signal correlations >99.99%, worst-case THD of -53.03 dBc, and pulse wave correlation >99.68% in frequency-domain with maximum NMSE% of 6e-6%; and 22.3-hour operation (3.3 Ah battery @ 150 mA). CONCLUSION: The system enables high-fidelity, power-efficient acquisition with validated signal integrity and adaptable multi-channel acquisition, addressing gaps in portable biosensing.