1999/09/01 by D.K. Jones, Derek K. Jones, M.A. Horsfield +3 · 7 citations
Medicine · #Advanced MRI Techniques and Applications #Advanced Neuroimaging Techniques and Applications #MRI in cancer diagnosis
paper · pdf · doi:10.1002/(sici)1522-2594(199909)42:3<515::aid-mrm14>3.0.co;2-q
openalex publication_date 1999/09/01 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/29
The optimization of acquisition parameters for precise measurement of diffusion in anisotropic systems is described. First, an algorithm is presented that minimizes the bias inherent in making measurements with a fixed set of gradient vector directions by spreading out measurements in 3-dimensional gradient vector space. Next, it is shown how the set of b-matrices and echo time can be optimized for estimating the diffusion tensor and its scalar invariants. The standard deviation in the estimate of the tensor trace in a water phantom was reduced by more than 40% and the artefactual anisotropy was reduced by more than 60% when using the optimized scheme compared with a more conventional scheme for the same scan time, and marked improvements are demonstrated in the human brain with the optimized sequences. Use of these optimal schemes results in reduced scan times, increased precision, or improved resolution in diffusion tensor images. Magn Reson Med 42:515-525, 1999.