First Clinical‐Scale Use of a Percutaneous EPR Oxygen Probe ( OxyTrack ) for Deep‐Tissue pO 2 Measurements in a Canine Cancer Patient

2026/05/30 by Conner S. Ubert, Maciej M. Kmiec, Sergey Petryakov +5
Agricultural and Biological Sciences · Biochemistry, Genetics and Molecular Biology · Medicine · #Electron Spin Resonance Studies #Photodynamic Therapy Research Studies #Radiation Effects and Dosimetry

paper · doi:10.1002/mrm.70449

openalex publication_date 2026/05/30 · openalex created_date 2026/05/31 · openalex updated_date 2026/08/01

Abstract

ABSTRACT Purpose To assess the safety, feasibility, and performance of a percutaneously insertable L‐band electron paramagnetic resonance (EPR) oxygen sensor (OxyTrack) for depth‐resolved pO 2 measurement in a spontaneous canine osteosarcoma, and to evaluate its translational potential. Methods A custom 1.15 GHz continuous‐wave EPR spectrometer was used together with two needle‐type resonators. An 8‐cm, 1‐mm‐OD OxyTrack measured pO 2 at five depths (2–38 mm) by incremental retraction; a 5‐cm, 0.5‐mm‐OD OxyTrack, mounted on a guiding cannula, enabled alternating measurements in tumor versus adjacent muscle at ∼30‐s intervals for 6 min. Data were analyzed in RStudio; means ± SEM were taken from the final minute of each depth. Results The pO 2 values (± SEM) in the osteosarcoma tumor were 49.26 ± 0.71 mmHg (at 2 mm depth), 50.82 ± 0.25 mmHg (at 8 mm), 19.62 ± 0.47 mmHg (at 18 mm), 4.78 ± 0.05 mmHg (at 28 mm), and 18.15 ± 0.30 mmHg (at 38 mm). The 28‐mm site exhibited severe hypoxia. Equilibration followed a mono‐exponential model ( R 2 > 0.98) with half‐times 0.35–1.17 min. Simultaneous muscle versus tumor measurements gave mean pO 2 of 33.9 ± 2.1 mmHg and 8.0 ± 0.7 mmHg, respectively (ratio ∼4.2:1). No adverse events occurred. Conclusion OxyTrack provides safe, real‐time, quantitative pO 2 assessment at depths > 3 cm, revealing marked intra‐tumoral oxygen heterogeneity that single‐point methods miss. These findings demonstrate feasibility for clinical translation of EPR oximetry to guide radiotherapy and hypoxia‐targeted therapies and support further translational development.

Related