Accuracy of Real-Time Echo-Planar Imaging Phase Contrast MRI

2022/07/28 by Liu, Pan, Fall, Sidy, Baledent, Olivier
#FOS: Electrical engineering #Signal Processing (eess.SP) #electronic engineering #information engineering

paper · doi:10.48550/arxiv.2207.13950

Abstract

Compared with CINE phase contrast MRI (CINE-PC), echo-planar imaging phase contrast (EPI-PC) can achieve realtime quantification of blood flow, with lower SNR. In this study, the pulsating real model of the simulated cerebral vasculature was used to verify the accuracy of EPI-PC. The imaging time of EPI-PC was 62ms/image at 100*60 spatial resolution. The reconstructed EPI-PC flow curve was extracted by homemade post-processing software. After comparison with the CINE-PC flow curve, it was concluded that EPI-PC can provide an average flow with less than 3% error, and its flow curve will be similar to the CINE-PC flow curve in shape.