vix.ing · top · new · best · stats · spec

Preliminary Design of a FADC Readout System for the Alpha/Beta Discrimination in a Large Area Plastic Scintillation Detector

2020/10/27 by Jingjun Wen, Daowei Dou, Wen, Jingjun +13
Agricultural and Biological Sciences · Engineering · Physics and Astronomy · #Advanced Chemical Sensor Technologies #FOS: Physical sciences #Instrumentation and Detectors (physics.ins-det) #Particle Detector Development and Performance #Radiation Effects and Dosimetry

paper · pdf · doi:10.48550/arxiv.2010.14093

openalex publication_date 2020/10/27 · openalex created_date 2020/11/09 · openalex updated_date 2026/07/28

Abstract

This paper describes a FADC Readout system developed for the tap water α/β dose monitoring system which is based on EJ444 phoswich scintillation detector and wavelength shifting fiber readout. The Readout system contains dual sampling channels that can supply sampling rate up to 1 GSPS, 14-Bit vertical resolution and adequate effective number of bits (9.7 Bits at 10 MHz), which is optimum for the discrimination of the minimal difference between alpha/beta signals. Moreover, the system is based on a ZYNQ SoC which provides high data throughput speed, low latency and excellent flexibility. As for the discrimination algorithms, a simple least-square classification method is used to discriminate the α/β signals and shows high discrimination capability. Besides, the front-end electronics and high voltage supply modules are also briefly introduced in the paper.