Measurements of a LYSO crystal array from threshold to 100 MeV

2024/09/23 by O. Beesley, Jill Carlton, Beesley, O. +35
Physics and Astronomy · #Crystallography and Radiation Phenomena #FOS: Physical sciences #High Energy Physics - Experiment (hep-ex) #Instrumentation and Detectors (physics.ins-det) #Nuclear Experiment (nucl-ex) #Nuclear Physics and Applications #Radiation Detection and Scintillator Technologies

paper · pdf · doi:10.48550/arxiv.2409.14691

openalex publication_date 2024/09/23 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/28

Abstract

We report measurements of ten custom-made high-homogeneity LYSO crystals. The investigation is motivated by the need for a compact, high-resolution, and fast electromagnetic calorimeter for a new rare pion decay experiment. Each 2.5× 2.5 × 18 cm3 crystal was first characterized for general light yield properties and then its longitudinal response uniformity and energy resolution were measured using low-energy gamma sources. The ten crystals were assembled as an array and subjected to a 30 - 100 MeV positron beam with excellent momentum definition. The energy and timing resolutions were measured as a function of energy, and the spatial resolution was determined at 70 MeV. An additional measurement using monoenergetic 17.6 MeV gammas produced through a p-Li resonance was later made after the photosensors used in positron testing were improved. As an example of the results, the energy resolution at 70 MeV of 1.52 ± 0.03% is more than two times better than reported results using previous generation LYSO crystals.