vix.ing · top · new · best · stats · spec

Estimation of Blood Glucose Level of Type-2 Diabetes Patients using Smartphone Video

2019/11/28 by Tauseef Tasin Chowdhury, Chowdhury, Tauseef Tasin, Tahmin Mishma +5
Engineering · Health Professions · Medicine · #Artificial Intelligence in Healthcare #ECG Monitoring and Analysis #FOS: Electrical engineering #Non-Invasive Vital Sign Monitoring #Signal Processing (eess.SP) #electronic engineering #information engineering

paper · pdf · doi:10.48550/arxiv.1911.12619

openalex publication_date 2019/11/28 · openalex created_date 2019/12/05 · openalex updated_date 2026/07/28

Abstract

This work proposes a smartphone video-based approach for the estimation of blood glucose in a non-invasive way. Videos using smartphone camera are collected from the tip of the subjects finger and the frames are subsequently converted into Photoplethysmography (PPG) waveform. Gaussian filter along with Asymmetric Least Square methods have been applied on the PPG signals to remove the high-frequency noise, optical and motion interferences. Different signal features such as Systolic and Diastolic Peaks, the time difference between consecutive peaks (DelT), First Derivative peaks, and Second derivative peaks etc have been extracted from the processed signal. Finally, Principal Component Regression (PCR) has been applied for the prediction of glucose level from the extracted features. The proposed model, while applied to an unbiased dataset, could predict the glucose level with a Standard Error of Prediction (SEP) of around 18.31 mg/dL.

Related