vix.ing · top · new · best · stats · spec

Response of germanium detectors for high-energy γ-rays by 27Al(p, γ)28Si at Ep=992 keV

2023/01/24 by Rurie Mizuno, Mizuno, Rurie, M. Nııkura +13
Physics and Astronomy · #FOS: Physical sciences #Instrumentation and Detectors (physics.ins-det) #Nuclear Experiment (nucl-ex) #Nuclear Physics and Applications #Nuclear physics research studies #X-ray Spectroscopy and Fluorescence Analysis

paper · pdf · doi:10.48550/arxiv.2301.09888

openalex publication_date 2023/01/24 · openalex created_date 2023/01/26 · openalex updated_date 2026/07/28

Abstract

The performance of germanium detectors for high-energy γ-rays was evaluated using a 992-keV resonance in the 27Al(p, γ)27Si reaction. The measurement was conducted at the RIKEN tandem accelerator. The energy of the excited state from the resonance was evaluated as 12540.7(2) keV. Using newly evaluated excitation energy, an energy calibration function and the photo-peak efficiency of Ge detectors up to 10.8-MeV photon were deduced. The energy accuracy is achieved at 0.3 keV for the overall energy region. This reaction provides reliable energy and efficiency standards for high-energy γ rays.

Related