An InGrid based Low Energy X-ray Detector

2014/10/01 by C. Krieger, Klaus Desch, Krieger, Christoph +7 · 1 citation
Physics and Astronomy · #Dark Matter and Cosmic Phenomena #FOS: Physical sciences #Instrumentation and Detectors (physics.ins-det) #Particle Detector Development and Performance #Radiation Detection and Scintillator Technologies

paper · doi:10.48550/arxiv.1410.0264

openalex publication_date 2014/10/01 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/29

Abstract

An X-ray detector based on the combination of an integrated Micromegas stage with a pixel chip has been built in order to be installed at the CERN Axion Solar Telescope. Due to its high granularity and spatial resolution this detector allows for a topological background suppression along with a detection threshold below 1 keV. Tests at the CAST Detector Lab show the detector's ability to detect X-ray photons down to an energy as low as 277 eV. The first background data taken after the installation at the CAST experiment underline the detector's performance with an average background rate of 5×10-5 /keV/cm2/s between 2 and 10 keV when using a lead shielding.

Cited by

Related