vix.ing · top · new · best · stats · spec

Mass testing of SiPMs for the CMVD at IICHEP

2022/08/24 by Mamta Jangra, Jangra, Mamta, Raj Bhupen +30
Physics and Astronomy · #Astrophysics and Cosmic Phenomena #FOS: Physical sciences #High Energy Physics - Experiment (hep-ex) #Instrumentation and Detectors (physics.ins-det) #Neutrino Physics Research #Particle Detector Development and Performance

paper · pdf · doi:10.48550/arxiv.2208.11446

openalex publication_date 2022/08/24 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/28

Abstract

A Cosmic Muon Veto Detector (CMVD) is being built around the mini-Iron Calorimeter (mini-ICAL) detector at the transit campus of the India based Neutrino Observatory, Madurai. The CMV detector will be made using extruded plastic scintillators with embedded wavelength shifting (WLS) fibres which propagate re-emitted photons of longer wavelengths to silicon photo-multipliers (SiPMs). The SiPMs detect these scintillation photons, producing electronic signals. The design goal for the cosmic muon veto efficiency of the CMV is >99.99% and fake veto rate less than 10-5. A testing system was developed, using an LED driver, to measure the noise rate and gain of each SiPM, and thus determine its overvoltage (Vov). This paper describes the test results and the analysed characteristics of about 3.5k SiPMs.

Related