Correcting directional dark-field x-ray imaging artefacts using position-dependent image deblurring and attenuation removal

2024/04/26 by Croughan, Michelle K, Paganin, David M, Alloo, Samantha J +4
#FOS: Physical sciences #Medical Physics (physics.med-ph) #Optics (physics.optics)

paper · doi:10.48550/arxiv.2404.17612

Abstract

In recent years, a novel x-ray imaging modality has emerged that reveals unresolved sample microstructure via a "dark-field image", which provides complementary information to conventional "bright-field" images, such as attenuation and phase-contrast modalities. This x-ray dark-field signal is produced by unresolved microstructures scattering the x-ray beam resulting in localised image blur. Dark-field retrieval techniques extract this blur to reconstruct a dark-field image. Unfortunately, the presence of non-dark-field blur such as source-size blur or the detector point-spread-function can affect the dark-field retrieval as they also blur the experimental image. In addition, dark-field images can be degraded by the artefacts induced by large intensity gradients from attenuation and propagation-based phase contrast, particularly around sample edges. By measuring any non-dark-field blurring across the image plane and removing it from experimental images, as well as removing attenuation and propagation-based phase contrast, we show that a directional dark-field image can be retrieved with fewer artefacts and more consistent quantitative measures. We present the details of these corrections and provide "before and after" directional dark-field images of samples imaged at a synchrotron source. This paper utilises single-grid directional dark-field imaging, but these corrections have the potential to be broadly applied to other x-ray imaging techniques.

Related