vix.ing · top · new · best · stats · spec

Comparison of Fast Amplifiers for Diamond Detectors

2013/10/03 by M. Osipenko, S. Minutoli, Osipenko, M. +22
Engineering · Materials Science · Medicine · Physics and Astronomy · #Advanced X-ray and CT Imaging #Diamond and Carbon-based Materials Research #FOS: Physical sciences #High Energy Physics - Experiment (hep-ex) #Instrumentation and Detectors (physics.ins-det) #Medical Imaging Techniques and Applications #Nuclear Experiment (nucl-ex) #hep-ex #nucl-ex #physics.ins-det

paper · pdf · doi:10.48550/arxiv.1310.1000

21 pages, 15 figures

arxiv created 2013/10/03 · openalex publication_date 2013/10/03 · arxiv updated 2013/10/22 · openalex created_date 2025/10/10 · openalex updated_date 2026/07/28

Abstract

The development of Chemical Vapour Deposition (CVD) diamond detectors requests for novel signal amplifiers, capable to match the superb signal-to-noise ratio and timing response of these detectors. Existing amplifiers are still far away from this goal and are the dominant contributors to the overall system noise and the main source of degradation of the energy and timing resolution. We tested a number of commercial amplifiers designed for diamond detector readout to identify the best solution for a particular application. This application required a deposited energy threshold below 100 keV and timing resolution of the order of 200 ps at 200 keV. None of tested amplifiers satisfies these requirements. The best solution to such application found to be the Cividec C6 amplifier, which allows 100 keV minimal threshold, but its coincidence timing resolution at 200 keV is as large as 1.2 ns.