2011/03/22 by Gou, Quanbu, Guo, Yiqing, Liu, Cheng +2
#FOS: Physical sciences #High Energy Physics - Experiment (hep-ex) #Instrumentation and Detectors (physics.ins-det)
paper · doi:10.48550/arxiv.1103.4213
This work aims at online calibrating the signal attenuation of the long cable used in Tibet Water Cherenkov Muon detector array-A (Tibet MD-A) under the 37 000 m2 Tibet air shower array. Based on the waterproof connection of the PMT (R360006) with the signal cable, and the characters of the high voltage divider, terminal reflection method is explored and used for measuring the signal attenuation with a practical way to eliminate the contribution of the pulse baseline (pedestal). Comparison measurement between it and QDC (Charge-to-Digital Conversion) data taking method are carried out by using open-ended cables, confirming that terminal reflection method is a fast and convenient method, and suitable to online calibrate the signal attenuation for Tibet MD-A. At 26° C, the measured attenuation coefficient with the 250-m cable, permanently connected to the PMT, is ~13.9%. Also, the cable frequency response is measured by using the sinusoidal signals, which could be used to study the time dispersion of the real signal produced with Tibet MD-A detector via Fourier analysis.